TM 11-6625-828-45
c. Procedure
Step
Control settings
No.
Test Equipment
Equipment under test
Test procedure
Performance standard
a. Adjust test set RF LEVEL SET
a.
Between 0. and 1.2 volts.
1
AN/URM-145:
MODULATION switch- at 30
Range to 3 volts
control for LEVEL SET mark
ure voltage with Ai/URM-14.).
b. Rotate test set RF LEVEL SET
Greater than 1.3 volts.
b.
control fully clockwise. Measure
voltage with AN/URM-145.
c. Rotate test set RF LEVEL SET
control fully counterclockwise.
Measure voltage with
AN/URM-145.
51 ohms 1.
2
Adjust RF LEVEL SET control for
a.
TS352B/U:
LEVEL SET mark on test set RF
termination cap.
FUNCTION: OHMS
meter
RANGE SWITCH
b.
RF meter indication should return
b. Remove AN/URM-145 from RF
OUTPUT, 1 VOLT CONNEC
to LEVEL SET mark within
1/32 inch.
TOR and replace termination
cap.
4-10