TM 11 5895-1096-40
Table 3-6. Encode Generator Circuit Card Troubleshooting
Malfunction
Probable Cause
Corrective Action
1. No CODE OUT or SYNC OUT signals generated.
a. Defective DC supply components
CR3, CR4, C14.
nal at + side of C14.
(1) If CORRECT, go to b below.
(2) If INCORRECT, disconnect test set from 120
VAC source. Using Multimeter AN/USM-223,
check CR3, CR4, C14 and replace defective
component.
b. Defective DC supply components
b. Disconnect test set from 120 VAC source.
CR2, C12, C13, R20.
C13, R20. Replace defective component.
2. Erratic of no CODE OUT signal generated.
a. Defective clock component.
SYNC OUR normal.
Oscilloscope OS-262P/U, connect CHI probe to
TP4. Adjust horizontal, vertical position, and trig-
NOTE
Only upper trace will be displayed on
oscilloscope.
(1) If CORRECT, go to b below.
(2) If INCORRECT, check Ul, U2, Yl and asso-
ciated components. Replace defective
component.
b. Defective divider U5B.
(1) If CORRECT, go to c below.
(2) In INCORRECT, replace U5.
c. Defective divider U5A:
c. Using Oscillscope OS-262P/U, connect CH2 probe
(1) If CORRECT, go to d below.
(2) If INCORRECT replace U5.
d. Defective reset line component.
d. Using Oscilloscope OS-262P/U, connect CH2
probe to TP3 and connect CHI probe to PI-20
(SYNC). Set test set MODULATION to MANUAL
and adjust oscilloscope controls as required to ob-
test set MODULATION to CONTINOUS and ob-
(1) If CORRECT, go to e below.
(2) If INCORRECT, check CRI, C8, RI5, R16, U6.
Replace defective component.
e. Defective code output component
U7A, U7B.
(1) If CORRECT, go to f below.
(2) If INCORRECT, replace U7.
f. Defective code output component
U 1.
(o). Replace UI.
3. Erratic or no SYNC OUT signal generated.
Defective sync component.
Using Oscilloscope OS-262P/U, connect CHI probe
CODE OUT normal.
to TP4, CH2 probe to U3-13. Adjust horizontal,
vertical position, and trigger level as required to ob-
(1) if INCORRECT, replace U3.
(2) if CORRECT, check R13, Q2. Replace defective
component.
4. Signal not present at 19.25 MHz OUT connector J4
a. Defective probe power supply.
a. Using Multimeter AN/USM-223, check LI, L2,
when using external probe amplifier assembly.
R18, R19, Q3, Q4. Replace defective component.
b. Defective DC blocking capacitor
b. Replace C9.
C9.
3-14