TM 11-6625-2632-14-1
s l e e v e s and aprons which the solvent
WARNING
cannot penetrate.
The fumes of TRICHLOROETHANE are
b. Remove grease or fungus from the exterior
toxic. Provide thorough ventilation
surfaces with a clean cloth dampened (not wet)
whenever it is used; avoid prolonged or
with trichloroethane.
repeated breathing of vapor. Do not use
near an o p e n f l a m e o r h o t s u r f a c e ;
3-8. Organizational Troubleshooting
trichloroethane is nonflammable but heat
Troubleshooting at the organizational category is
c o n v e r t s the fumes to a highly toxic
confined to a visual indication of possible or actual
phosgene gas the inhalation of which could
trouble and is based on the monthly preventive
result in serious injury or death. Prolonged
maintenance checks and equipment performance
or repeated skin contact with
during use. Any malfunction that is beyond the
trichloroethane can cause skin in-
scope of organizational maintenance to correct
flammation. When necessary, use gloves,
shall be referred to general support category
maintenance personnel.
Section
IV.
MAINTENANCE
storage position in the test set cover.
Organizational maintenance is limited to:
(3) place new cable assembly in its storage
a. Replacing defective or missing connector
position in the test set cover.
adapters.
(4) perform the procedures given in a(5) and
(6) above.
b . Replacing defective or missing cable
c. Spanner Wrench Replacement.
assemblies.
c. Replacing d e f e c t i v e or missing spanner
(1) Perform the procedures given in a(1) and
(2) above.
wrench.
d . Replacing defective or missing knob on
(2) Remove defective spanner wrench from its
storage position in the test set cover.
frequency selector switch.
(3) Place new spanner wrench in its storage
3-10. Replacement Procedures
position in the test set cover.
(4) Perform the procedures given in a(5) and
a. Connector Adapter Replacement.
(6) above.
(1) Open elbow catch securing test set cover.
d. Frequency Selector Switch Knob Replace-
(2) Open test set cover.
ment.
(3) Remove defective connector adapter from
(1) Perform the procedures given in a(1) and
its storage position in the test set cover.
(2) above.
(4) Place new connector adapter in its storage
(2) Loosen set screws and remove knob by
position in the test set cover.
pulling knob off of frequency selector switch shaft.
(5) Close test set cover.
(3) Reinstall new knob by pushing knob on to
(6) Secure test set cover with elbow catch.
frequency selector switch shaft and tighten the set
b. Cable Assembly Replacement.
screws.
(1) Perform the procedures given in a(1) and
(4) Perform the procedures given in a(5) and
(2) above.
(6) above.
(2) Remove defective cable assembly from its